beautypg.com

10 appendix: glossary, 10 appendix: glossary -1 – YSI IQ S ENSOR N ET NitraVis 70x IQ User Manual

Page 55

background image

NitraVis 70x IQ

Appendix: Glossary

10 - 1

ba75950e02

11/2012

10 Appendix: Glossary

Resolution

Smallest difference between two measured values that can be

displayed by a meter.

Absorbance

(Absorption)

Attenuation of the light beam when it penetrates the sample.

Adjusting

To manipulate a measuring system so that the relevant value (e. g. the

displayed value) differs as little as possible from the correct value or

a value that is regarded as correct, or that the difference remains

within the tolerance.

Calibration

Comparing the value from a measuring system (e. g. the displayed

value) to the correct value or a value that is regarded as correct.

Often, this expression is also used when the measuring system is

adjusted at the same time (see adjusting).

Calibration value pair

Value pair consisting of the raw value measured by the UV-VIS

sensor, and a reference value measured e. g. by a laboratory deter-

mination. The calibration value pairs are the result of the user calibra-

tion.

Concentration

Mass or amount of a dissolved substance per volume, e. g. in g/L or

mol/L.

Measuring system

The measuring system comprises all the devices used for measuring,

e. g. measuring instrument and sensor. In addition, there is the cable

and possibly an amplifier, terminal strip and armature.

Measured parameter

The measured parameter is the physical dimension determined by

measuring, e. g. pH, conductivity or D.O. concentration.

Measured parameter

(correlated)

With the measuring method used, a correlated measured parameter

is not determined directly. A correlated measured parameter can,

however, be derived from a determinable measured parameter. The

precondition for this is that there is a reproducible interrelationship

(correlation) between the determinable and correlated measured

parameter.

Test sample

Designation of the test sample ready to be measured.

Measuring gap

The measuring gap is between the two measurement windows. In the

measuring gap, the light beam penetrates the test sample.

Measured value

The measured value is the special value of a measured parameter to

be determined. It is given as a combination of the numerical value and

unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
The measured value is determined from the raw value taking into

account the data from a user calibration.

The measured value is displayed together with the raw value in the

measured value display.