beautypg.com

YSI IQ S ENSOR N ET NitraVis 70x IQ User Manual

Page 13

background image

NitraVis 70x IQ

Commissioning

3 - 3

ba75950e02

11/2012

Vertical installation

For vertical installation, mount the sensor in a position that allows inter-

fering elements in the measuring gap to escape with the current. In the

case of vertical installation, the minimum flow rate is 0.2 m/s so the

measuring gap is kept clean optimally.

Fig. 3-2 Vertical installation

Protection against shocks

When installing the sensor, always make sure it cannot knock against

a wall or another obstacle. Make sure the safety margins are sufficient

(e. g. with suspended mounting).

Mount the enclosed shock protectors (see section 3.3.2). Their soft rub-

ber compound absorbs hard impacts.

Fig. 3-3 Sensor with mounted shock protectors

Flow direction

easuring gap opening