YSI IQ S ENSOR N ET NitraVis 70x IQ User Manual
Page 33
NitraVis 70x IQ
Measuring / Operation
4 - 13
ba75950e02
11/2012
If no valid zero adjustment is possible under the current conditions, you
can restore the readiness for measurement by measuring with an older
valid zero adjustment.
To do so, select the
Zero adjustment
setting in the sensor settings
Factory
(factory zero adjustment) or
User
(last valid user zero adjust-
ment).
Then check the measurement results with a
Sensor check (standard)
.
Suitable standards (check equipment) for your sensor are available as
accessories (see chapter 6).
Result
after
Sensor check (standard)
Completing the sensor
check or
Zero adjustment
After a successful sensor check or
Zero adjustment
, return to measure-
ment as follows:
Store the sensor sleeve in a clean environment, e. g. in a plastic bag,
until the next sensor check or
Zero adjustment
.
Result
Meaning
OK
Sensor check (standard)
successful.
Error
Sensor check (standard)
not successful.
Possible cause: wrong or expired standard
17.
Confirm the calibration result with
18.
Confirm the message,
"After successful cal ..."
with
The display returns to the measured value display. The mea-
sured value flashes because the sensor is still in the mainte-
nance condition.
19.
Pour away the ultrapure water or standard solution and
remove the sensor sleeve.
20.
Screw in the blind plug or compressed air connector and install
the sensor at the measuring site.
21.
After the
Sensor check (standard)
:
Thoroughly rinse the sensor and sensor sleeve with ultrapure
water.
22.
Switch off the maintenance condition (use <> to select
the UV-VIS sensor, press
Display/Options
menu).
23.
Switch to the measured value display with
value stops flashing.