Measuring / Operation

NitraVis 70x IQ

4 - 12

ba75950e02

11/2012

Up to this point, you can break off the calibration procedure at any time

with the key. The system continues to work with the old calibra-

tion data. The maintenance condition, however, has to be switched off

again in any case.

After starting the calibration with the key (step 16.) you can no

longer cancel the calibration procedure.

Result

after

Sensor check

(H2O dest.)

If the rating number is not in the zero range:

Repeat the cleaning procedure and the

Sensor check (H2O dest.)

until

the rating number no longer changes significantly.

If the rating number is still outside the zero range, a new zero adjust-

ment is necessary.

Result

after

Zero

adjustment

If the result is erroneous, clean the sensor and all equipment once

again. Then carry out a new

Zero adjustment

and when doing so make

sure the conditions are absolutely clean.

15.

Confirm each checklist with until the

Start measurement

display appears.

16.

Confirm with .

The measurement is started. After the end of the measurement

the result appears. With the

Zero adjustment

, the result is

entered in the calibration history at the same time.

Rating number

Meaning

Zero range

(-40 ... +40)

Sensor check (H2O dest.)

successful (OK)

(0 = no deviation)

Positive number

(+40 ... +1000)

Measured values were too low.

Possible cause:

The measurement windows are contaminated.

Negative number

(-40 ... -1000)

Measured values were too high.

Possible cause:

During the last zero adjustment, the measure-

ment windows were contaminated more than dur-

ing the current sensor check.

Result

Meaning

Sensor released for

measurement

OK

Zero adjustment

successful.

Yes

Error

Zero adjustment

not success-

ful.

No ("----")