YSI IQ S ENSOR N ET NitraVis 70x IQ User Manual
Page 32
Measuring / Operation
NitraVis 70x IQ
4 - 12
ba75950e02
11/2012
Up to this point, you can break off the calibration procedure at any time
with the
tion data. The maintenance condition, however, has to be switched off
again in any case.
After starting the calibration with the
longer cancel the calibration procedure.
Result
after
Sensor check
(H2O dest.)
If the rating number is not in the zero range:
Repeat the cleaning procedure and the
Sensor check (H2O dest.)
until
the rating number no longer changes significantly.
If the rating number is still outside the zero range, a new zero adjust-
ment is necessary.
Result
after
Zero
adjustment
If the result is erroneous, clean the sensor and all equipment once
again. Then carry out a new
Zero adjustment
and when doing so make
sure the conditions are absolutely clean.
15.
Confirm each checklist with
Start measurement
display appears.
16.
Confirm with
The measurement is started. After the end of the measurement
the result appears. With the
Zero adjustment
, the result is
entered in the calibration history at the same time.
Rating number
Meaning
Zero range
(-40 ... +40)
Sensor check (H2O dest.)
successful (OK)
(0 = no deviation)
Positive number
(+40 ... +1000)
Measured values were too low.
Possible cause:
The measurement windows are contaminated.
Negative number
(-40 ... -1000)
Measured values were too high.
Possible cause:
During the last zero adjustment, the measure-
ment windows were contaminated more than dur-
ing the current sensor check.
Result
Meaning
Sensor released for
measurement
OK
Zero adjustment
successful.
Yes
Error
Zero adjustment
not success-
ful.
No ("----")