beautypg.com

4 initial commissioning, 1 general information, 4 initial commissioning -7 – YSI IQ S ENSOR N ET NitraVis 70x IQ User Manual

Page 17: General information -7

background image

NitraVis 70x IQ

Commissioning

3 - 7

ba75950e02

11/2012

3.4

Initial commissioning

3.4.1 General information

CAUTION

Never look into the measuring window or put objects into the measur-

ing window during operation! The emitted UV radiation can damage the

eyes!

In the case of authorized use, inadvertent eye contact with the light

beam is not possible.

Measuring location

The NitraVis 701 IQ sensor is prepared for operation in the aeration

tank and calibrated in the factory. For use in the inlet you only have to

change the

Measuring location

setting.

The NitraVis 705 IQ sensor is prepared for operation in the outlet of the

waste water treatment plant and calibrated in the factory.
The sensor is immediately ready to measure.
How to adjust the settings is described in the following section (section

3.4.2).

User calibration after initial

commissioning

On the basis of a reference measurement ("laboratory value"), check

whether the default settings deliver sufficiently accurate measured

values or whether you want to carry out a user calibration (see section

4.3 C

ALIBRATION

).