3 calibration, 1 overview, 3 calibration -3 – YSI IQ S ENSOR N ET NitraVis 70x IQ User Manual
Page 23: Overview -3, Section, Alibration
NitraVis 70x IQ
Measuring / Operation
4 - 3
ba75950e02
11/2012
4.3
Calibration
4.3.1 Overview
In the delivery condition, the sensor is precalibrated and ready to mea-
sure (see section 3.4.1).
There are two types of calibration that are included into the calculation
of the measured value at different points (see section 4.1):
User calibration (see section 4.3.2)
Sensor check/
Zero adjustment
An overview diagram of the calibration procedures can be found on
page 4 - 9.
User calibration
With a user calibration you compensate for the influence of the test
sample. The user calibration requires a successful
Sensor check (H2O
dest.)
and a valid
Zero adjustment
.
We recommend to check the current user calibration against indepen-
dent comparison measurements if you suspect a basic change of the
characteristics of the test sample. If the comparison measurements
show clear deviations, these can be compensated by a user calibration.
Recalibration after
changing the measuring
location
The data of the user calibration are stored in the IQ S
ENSOR
N
ET
con-
troller and thus assigned to the measuring location (not to the sensor).
The calibration data of the
Zero adjustment
are stored in the sensor.
Each time after the application or measurement location
(and the setting,
Measuring location
) was changed, we
recommend to carry out a
Sensor check (H2O dest.)
(and, if the
Sensor check (H2O dest.)
was not successful,
an additional
Zero adjustment
if necessary) with a subse-
quent user calibration.