2 info messages, Info messages -3 – Xylem IQ SensorNet FDO 70x IQ (SW) User Manual
Page 45

FDO
®
70x IQ (SW)
Indexes
9 - 3
ba76014e01
02/2012
9.1.2
Info messages
Message code
Message text
IC133x
Sensor has been successfully calibrated
* For calibration data, see calibration history
IC333x
Factory calibration has been activated. Make sure the sensor operates
correctly.
IC433x
Last valid user calibration has been activated. Make sure the sensor
operates correctly.
IC533x
Invalid user calibration has been replaced by last valid user calibration.
Caution! Wrong measured values possible. Carry out a new successful
user calibration to make sure the sensor operates correctlyes correctly.
IC633x
The invalid user calibration was replaced by the factory calibration.
Caution! Wrong measured values possible. Check whether the sensor
operates correctly or carry out a new successful user calibration user
calibration.
IC733x
This message displays the result of the last check (see section 4.2.2)
IS133x
This message displays the information stored in the sensor cap
(see section 1.2)