EC833x
Sensor could not be calibrated,
sensor blocked for measurement
Cause: instable signal
* Check temperature adjustment
* Check calibration conditions (see operating manual)
* Repeat calibration
EC933x
Calibration error, measurement disabled
Cause: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor according to
op.instructions
* Screw on sensor cap securely
* Repeat calibration
* Replace sensor cap
EI133x
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components
EI233x
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components,
replace components
ES133x
Component hardware defective
* Contact service