What to do if...

FDO

®

70x IQ (SW)

6 - 2

ba76014e01

02/2012

Measured value invalid

( "----" displayed)

Incorrect temperature

display

Cause

Remedy

– User calibration unsuccessful.

The sensor is blocked for mea-
surement.

– As a temporary measure to

quickly restore the readiness
for service: Activate the fac-
tory calibration (see section
4.2.5)

– For exact measurements,

carry out a function check and
repeat the user calibration if
necessary.

Cause

Remedy

– Temperature sensor defective

– Return the sensor