FDO
®
70x IQ (SW)
Technical data
7 - 1
ba76014e01
02/2012
7
Technical data
7.1
General measurement characteristics
Note
The measurement characteristics are primarily determined by the sen-
sor cap type. The relevant data are given in the sections 7.5 and 7.6.
Measuring principle
Optical measurement based on photoluminescence.
Measurement in water
According to solubility function according to ISO 5814
Measurement in salt-
contaminated
wastewater
Salinity input from 2.0 ... 70.0;
corresponds to 3.4 mS/cm ... 86.2 mS/cm at T
REF
20 °C
(salinity measurement according to IOT = International Oceanographic
Tables)
Temperature
measurement
Temperature
compensation
In the range -5 °C ... + 50 °C (23 ... 122 °F)
7.2
Application conditions
Pressure resistance
Sensor with connected SACIQ (SW) sensor connection cable:
The sensor meets all requirements according to article 3(3) of 97/23/
EC ("pressure equipment directive").
Type of protection
Sensor with connected SACIQ (SW) sensor connection cable:
IP 68, 10 bar (10
6
Pa)
Immersion depth
Min. 10 cm; max. 100 m depth
Operating position
Any
Temperature sensor
Integrated NTC
Measuring range
- 5 °C ... + 50 °C (23 ... 122 °F)
Accuracy
± 0.5 K
Resolution
0.1 K
Max. allowed overpressure
10
6
Pa (10 bar)