Appendix a - specifications – YSI 3200 User Manual
Page 71

APPENDIX A - SPECIFICATIONS
Modes
Conductivity
Conductance
Resistivity
Resistance
Salinity
TDS
Temperature
Conductance Range
Accuracy
Resolution
Frequency
0 - 0.9999
µS
± 0.30% full scale
0.0001
µS
40 Hz
0.950 - 9.999
µS
± 0.20% full scale
0.001
µS
80 Hz
9.50
µS - 99.99 µS
± 0.10% full scale
0.01
µS
290 Hz
95.0 - 999.9
µS
± 0.10% full scale
0.1
µS
1010 Hz
950 - 9999
µS
± 0.10% full scale
1
µS
1010 Hz
9.50 - 99.99 mS
± 0.10% full scale
0.01 mS
1010 Hz
95.0 - 999.9 mS
± 0.30% full scale
0.1 mS
1460 Hz
0.95 - 3.00 S
± 1.0% full scale
0.01 S
1460 Hz
Resistance Range
Accuracy Resolution
Frequency
0 - 9.999
Ω
± 0.2% full scale
0.001
Ω
1460 Hz
0 - 99.99
Ω
± 0.1% full scale
0.01
Ω
1460 Hz
0 - 999.9
Ω
± 0.1% full scale
0.1
Ω
1010 Hz
0 - 9.999 k
Ω
± 0.1% full scale
0.001 k
Ω
1010 Hz
0 - 99.99 k
Ω
± 0.1% full scale
0.01 k
Ω
1010 Hz
100.0 - 999.9 k
Ω
± 0.2% full scale
0.1 k
Ω
290 Hz
1.00 - 9.99 M
Ω
± 0.5% full scale
0.01 M
Ω
80 Hz
10.0 - 29.9 M
Ω
± 1% full scale
0.1 M
Ω
40 Hz
Salinity Range Accuracy
Resolution
0-80 ppt (NaCl)
±.1 ppt
0.1 ppt
TDS Range
Accuracy
Resolution
0-19,999 mg/L
0.50%
1 mg/L
Temperature Range
Accuracy Resolution
-5 - 100°C
±0.1°C
0.01°C
YSI Incorporated
Model 3200
67