999us – YSI 3200 User Manual
Page 28

System Configuration and Operation
Section 3
13.2uS
K=0.100
25.0°C
11:34
AM
24.38°
C
1.50%/°C
T CELL
CURRENT RANGE [9.50 - 99.99uS]
CELL
SETUP
DIAGNOSTICS
RANGE
OPERATION
If the range selected cannot display the current reading, an “OUT OF RANGE” message is
displayed as shown below.
9.999uS
K=0.100
25.0°C
11:34
AM
24.38°
C
1.50%/°C
T CELL
OUT OF RANGE
CELL
SETUP
DIAGNOSTICS
RANGE
OPERATION
Each time the [RANGE] soft key is pressed, the 3200 hardware will change the current
conductance/resistance range. If measurements are currently displayed as conductance or
resistance, you will see the range change on the display. If, however, you are displaying
conductivity, resistivity, TDS or salinity, as you change ranges, the display will automatically
select the best resolution and the hardware range change will not be visible.
Example:
With the 3200 in manual range mode, 9.50 - 99.99mS range, the display is as follows:
1017
uS/cm
10.27
mS
OPERATION
984.6
Ωcm
97.34
Ω
K=0.098
25.0°C
691
mg/L
TDS
0.5
ppt
SAL
1.91%/°C
T CELL
CELL
SETUP
DIAG
RANGE
If the [RANGE] soft key is pressed, the hardware switches to the 95.0mS - 2.99S range. The
conductance and resistance fields reflect the change, but the conductivity, resistivity, TDS and
salinity fields remain essentially unchanged.
YSI Incorporated
Model 3200
24