Introduction
Section
1
1.2 USER
FEATURES
• Audio feedback
• Soft key/menu system
• Real time clock
• Auto linear temperature compensation calibration
Two temperature points are tested in a known solution to automatically
determine the solution’s compensation slope. Alternatively, the user may input a
temperature compensation factor directly (0 - 10%/°C).
• Auto non-linear temperature compensation calibration
This is a technique where a known solution is used to calibrate over a wide range of
temperature. The solution is ramped though a temperature range while the instrument
automatically and continuously generates calibration data. The calibration would be done
once and the results stored in non-volatile memory for recall any time that solution is
tested. Data for six solutions can be stored at any one time.
The user may also select from 2 pre-programmed non-linear temperature
compensation curves.
• Single or multi-point cell calibration
The cell constant can be calibrated at several points and stored in permanent
memory. During a reading, the closest calibration point is used as the cell
constant.
• High/low alarm