Operation

111

Single-point

calibration

Use a buffer solution in the range pH = 7.0 ± 0.5 for this pro-
cedure.

Note
Only the electrode asymmetry is determined in single-point
calibration. The slope of the last two-point calibration is
retained.

Note
Step 2 is not required if you use a temperature probe. The
TP message indicates an active temperature measurement.

1 Press the key repeatedly until

ASY

and the

ConCal

function display appear.

2 Set the temperature of the buffer solution by pressing

<

> <

>.

3 Submerse the pH electrode in the buffer solution.

4 Press the key.

The measured pH value appears on the display.

5 Set the nominal pH value of the buffer solution (at the

current temperature) by pressing the <

> <

> keys.

6 Press the key. The value of the asym-

metry (mV) and the sensor symbol for the evaluation
of the electrode appears on the display.

7 To return to the measuring mode: Press the

key.