beautypg.com

Xylem CG 843P User Manual

Page 13

background image

Overview

85

Electrodes

Socket acc. to DIN 19 262 or BNC
4-mm socket for reference electode

Temp. probe

4-mm socket for temperature probe

Interface/
analog output

Bidirectional RS 232 interface or ana-
log output with autom. recognition of
the PC, printer, or recorder connected

Plug-in power
supply

2 pole special Friwo

Connectors

AutoCal TEC

With Technical buffers, pH values
2.00; 4.00; 7.00; 10.00 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C

AutoCal DIN

With standard buffers according to
DIN 19 266/NIST, pH values:
1.68; 4.01; 6.87; 9.18 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C

ConCal

With any buffers; one and two-point
calibration, manual acceptance of
measured values and temp. input

Print principle

Matrix thermoprint

signs per line

26

printing width

46.1 mm

width of printing paper

58 mm

printing speed

40 signs/sec

Calibration modes

Printer