beautypg.com

1 measuring the ph value – Xylem CG 843P User Manual

Page 24

background image

Operation

96

4.2.1 Measuring the pH value

AutoRead AR

(Drift control)

The

AutoRead

function (drift control) checks the stability of

the measurement signal. The stability has a considerable
effect on the reproducibility of the measured values.

For identical measurement conditions, the following criteria
apply:

l

pH value: better than 0.02 (response time: > 30 s)

1 Perform the preparatory activities according to sec-

tion 4.2.

2 Immerse the pH electrode into the test sample.

3 Press the key until

pH

appears in the status

display. The pH value appears on the display.

1 Call up the pH measuring mode by pressing .

2 Activate the AutoRead function by pressing

read>.
The current measured value is frozen (Hold function).

3 Start the AutoRead function by pressing

.
AR flashes on the display until a stable measured val-
ue is reached. This measured value is transmitted to
the printer/RS interface.

4 If necessary, start the next AutoRead measurement

by pressing .