3 calibrating – Xylem CG 843P User Manual
Page 27

Operation
99
4.3
Calibrating
Why calibrate?
pH electrodes age. This changes the asymmetry (zero
point) and slope of the pH electrode. As a result, an inexact
measured value is displayed. Calibration determines the
current values of the asymmetry and slope of the electrode
and they are stored in the instrument.
Thus, you should calibrate at regular intervals.
When to calibrate?
l
After connecting another electrode
l
When the sensor symbol flashes:
– after expiry of the calibration interval
– after a voltage interruption, e.g. battery change
You can choose between 3 calibration procedures:
AutoCal TEC
is specially adapted to the Technical buffer solutions as a
fully automatic two or three-point calibration. The buffer
solutions are automatically recognized by the instrument.
The following pH values are available:
2.00, 4.00, 7.00, 10.00.
AutoCal DIN
is specially adapted to permanently programmed buffer
solutions according to DIN 19266 as a fully automatic two or
three-point calibration. The buffer solutions are automati-
cally recognized by the instrument.
The following pH values are available:
1.68, 4.01, 6.87, 9.18.
ConCal
is the conventional two-point calibration with 2 freely select-
able buffer solutions or single-point calibration as the rapid
method.