beautypg.com

Xylem CG 843P User Manual

Page 28

background image

Operation

100

AutoRead

In calibration using AutoCal TEC and AutoCal DIN, the

AutoRead

function is automatically activated.

The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
.

Calibration protocol

The calibration protocol contains the calibration data of the
current calibration. You can call up the calibration protocol
by outputting the data storage (section 4.4.3).

Note
You can automatically print out a calibration protocol after
the calibration. To do so, switch on the printer before the cal-
ibration (Print LED lights up green). After a valid calibration,
the protocol is printed.

Sample printout:

CALIBRATION PROTOCOL
02.03.99 14:19
Device No.: 12345678
CALIBRATION pH
Cal Time: 01.03.99 / 15:20
Cal Interval: 7d
AutoCal DIN Tauto
Buffer 1 1.679
Buffer 2 4.008 *
Buffer 3 6.865
Buffer 4 9.180 *
C1 174.1mV 25.0°C
C2 -133.3mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++