Page 34

background image

Operation

106

4.3.3 AutoCal DIN

Use two or three different DIN buffer solutions for this proce-
dure in increasing or decreasing order:

Note
Steps 2, 6 and 13 are not required if you use a temperature
probe.

DIN buffer solution

pH

A

1.68

C

4.01

D

6.87

F

9.18

1 Press the key repeatedly until

Cd1

and the

AutoCal DIN

function display appear.

2 If necessary, set the temperature of the buffer solu-

tion by pressing <

> <

>.

3 Submerse the pH electrode in the first buffer solution.

4 Press the key.

AR

flashes on the display.

The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,

Cd2

ap-

pears.