Advanced instruments inc – Analytical Industries GPR-2600 Series Oxygen Analyzer User Manual
Page 40

Advanced Instruments Inc.
40
MAIN MENU
Sample
* Span
Zero
Alarm
System
Standby
Sample
20.5 %
Auto Range
85⁰F 100Kpa
0 to 1 %
12/31/07 12:00:00
The following menu appears:
SPAN Sample
Factory Default
* Calibrate
Auto Range 0 to 1 %
85⁰F 100Kpa 12/31/07 12:00:00
Advance the cursor (*) to the “Manual Span” option and press the green ENTER key to accept the selection.
The following menu appears:
SPAN GAS Standby
Enter as PPM
Enter as PPB
* Enter as %
Auto Range 0 to 1 %
85⁰F 100Kpa 12/31/07 12:00:00
Enter span gas units appropriate to the numerical span gas value and press ENTER key.
Note: Select “Enter as %” for air calibration.