Range Mode Selection
Advance the cursor (*) to the “Sample” option as illustrated and press the green ENTER key to accept the selection. 
 
MAIN MENU
* Sample 
 Span 
 Zero 
 Alarm 
 System 
 Standby 
Standby
Auto Range 
85⁰F 100Kpa 
 
12/31/07 12:00:00 
 
The following menu appears: 
 
* SAMPLE
 Auto Range 
 Manual Range 
 Bypass 
 Standby 
 
 
Standby
Auto Range 
85⁰F 100Kpa 
 
12/31/07 12:00:00 
 
The analyzer is equipped with five (5) standard measuring ranges (see specification) and provides users with a choice of 
sampling modes. By accessing the MAIN MENU, users may select either the Auto Range or a fixed Manual Range) mode. 
 
Note:
For calibration purposes, use of the Auto Range mode is recommended. However, the user can select a fixed Manual 
Range (full scale) as dictated by the oxygen content of the calibration gas. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 