2: Calibration

6

JAZ-A-IRRAD-02-201011

8. Set the spectrometer for the optimal signal-to-noise ratio by clicking Set Automatically or

making the following individual selections:

Integration Time: The peak of the signal should be close to 55,704 counts without saturation
of the detector (peak > 64,000 counts).

Scans to Average: 30

Boxcar width: Leave setting at 5.

Click Next when settings have been selected.

9. Click

to take a reference measurement. This data will be compared with the calibrated data of

your light source. Click Next.