Performing a calibration using spectrasuite – Ocean Optics Jaz Absolute Irradiance User Manual
Page 10

2: Calibration
4
JAZ-A-IRRAD-02-201011
If you have a calibration file available (via SpectraSuite or SPEC-CAL) go on to Chapter 3:
.
If you are performing your own calibration, use the following procedure.
Performing a Calibration Using SpectraSuite
If you have purchased a calibration light source though Ocean Optics, you can use this lamp to create the
absolute irradiance calibration of the spectrometer and probe. The procedure below describes how to
acquire this calibration file.
►
Procedure
1. Connect the spectrometer to the fiber optics probe (fiber, cosine corrector or integrating sphere).
2. Connect the fiber optics probe to the calibration light source.
3. Turn on the light source. Allow for proper warming up of the light source as stated in the light
source’s manual.
4. Start SpectraSuite. The raw spectrometer data appears on the screen (Scope mode).
5. Navigate to File | New | New Absolute Irradiance Wizard to launch the Absolute Irradiance
wizard.
- Apex 785 Raman (1 page)
- STS-UV (2 pages)
- TR2 Engineering Note (4 pages)
- SAD500 Communications and Control (19 pages)
- Red Tide USB650 Install (26 pages)
- Fiber Optic Termination Kit (6 pages)
- Transmissive pH Probe (10 pages)
- Remora (42 pages)
- PlasCalc (59 pages)
- Correcting Device Driver Issues (8 pages)
- ecoVis Krypton Light Source (16 pages)
- LPC-500CM (28 pages)
- HPX-2000 (24 pages)
- ADC1000-USB (27 pages)
- Torus Operating Instructions (30 pages)
- ADC2000-PCI (13 pages)
- Sensors for Real-Time Analysis (2 pages)
- IDRaman reader (2 pages)
- DH-2000-CAL (30 pages)
- QE65 Pro (32 pages)
- Collimating Lenses (2 pages)
- D1000 (2 pages)
- IDRaman mini (2 pages)
- HR2000CG-UV-NIR (42 pages)
- Cool Red (1 page)
- HL-2000 (20 pages)
- XE-1 Xenon (6 pages)
- USB-ADP Serial Adapters (3 pages)
- CHEM2000 (36 pages)
- Deuterium-Halogen Calibration Light Source (19 pages)
- NeoFox Engineering Note (30 pages)
- QE Pro (2 pages)
- OOIBase32 (140 pages)
- AR-1 Argon (6 pages)
- pH Sensor Patches, Probes and Cuvettes (36 pages)
- SpecLine Offline Spectroscopy (60 pages)
- HL-2000-HP-232 (26 pages)
- External Triggering Options Instructions for Spectrometers with Firmware Version 3.0 and Above (16 pages)
- Breakout Box (10 pages)
- USB-ISS-UV_VIS (4 pages)
- ISS-UV_VIS (6 pages)
- OOIColor (14 pages)
- LS-1 Series (12 pages)
- Apex Install (24 pages)
- DH-2000 (34 pages)