Maya2000 series spectrometers specifications, Ccd detector specifications, Maya2000 series – Ocean Optics Maya2000Pro VIS-NIR Install User Manual
Page 26: Spectrometers specifications
![background image](/manuals/292533/26/background.png)
B: Specifications
20
020-00000-000-02-201207
Maya2000 Series Spectrometers Specifications
The following sections provide specification information for the CCD detector in the Maya2000 Series
spectrometers, as well as for each model Maya2000 Series Spectrometer itself.
CCD Detector Specifications
Specification
Maya2000
Maya2000 Pro
Maya200Pro VIS-NIR
Detector
Hamamatsu S9840,
back-thinned FFT-CCD
Hamamatsu S10420,
back-thinned FFT-CCD
Hamamatsu S11510,
back-thinned FFT-CCD
Thermoelectric
cooling
No
Number of pixels
All: 2080 x 20
Active: 2048 x 14
All: 2068 x 70
Active: 2048 x 64
Spectral range
200-1100 nm with window, Deep UV option available (150nm). Deep UV option
includes purge port and window removal.
175-1100 nm with HC1 grating.
Pixel size
14 µm square
Pixel well depth
130 Ke-
200 Ke-
Column height
196 µm square
896 µm square
Detector active area
(mm)
28.672 horizontal x 0.196
vertical
28.672 horizontal x 0.896 vertical
Quantum efficiency:
Peak QE
QE @ 250 nm
>90%
55%
75% at 600 nm
60%
85% at 700 nm