Measurement and power solutions, Models p0621p and n specifications – Trek P0621 High-Voltage Power Amplifier User Manual
Page 2

Features
High Voltage On/Off
Local
Individual push-button switches
Remote
TTL high turns OFF the high voltage; TTL low
turns on the high voltage
Dynamics
Adjustments
Graduated 1-turn potentiometer used to
optimize the AC response for various load
parameters
Current Limit/Trip
Switch selectable for limit or trip. Graduated 1-
turn potentiometer adjusts from 0 to 20 mA
Out of Regulation
LED illuminates and BNC provides a TTL low
when P0621 fails to produce HV output such as
during a current limit
Trip Status
LED illuminates and BNC provides a TTL low
when HV is disabled due to the output current
exceeding the current trip level, a high voltage
fault is detected or the top cover is removed
Fault Status
BNC provides a TTL low when P0621 is out of
regulation for greater than 500 ms
Mechanical
Dimensions
222 mm H x 432 mm W x 584 mm D
(8.75” H x 17” W x 23” D)
Weight
24.9 kg (55 lb)
HV Connector
Caton High Voltage Connector
BNC Connectors
Amplifier input, voltage monitor, current
monitor, remote HV ON/OFF, out of regulation,
fault/trip status
Operating Conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative Humidity
To 75%, noncondensing
Altitude
To 2000 meters (6561.68 ft.)
Electrical
Line Voltage
Factory Set for one of two ranges:
90 to 127 V AC or 180 to 250 V AC,
either at 48 to 63 Hz
AC Line Receptacle
Standard 3-prong with integral fuse holder
Power Consumption
1000 VA, maximum
Supplied Accessories
Operators’ Manual
PN: 23147
HV Output Cable
PN: 47067
Line Cord (for 104 to
127 V AC)
PN: N5011
Line Cord (for 180 to
250 V AC)
Selected per geographic destination
Copyright ã 2012 TREK, INC. All specifications are subject to change. 1251/JRB
Models P0621P and N Specifications
Performance
Output Voltage
Positive Polarity
(P0621P)
0 to +30 kV DC or peak AC
Output Voltage
Negative Polarity
(P0621N)
0 to -30 kV DC or peak AC
Output Current
0 to ±20 mA
Input Voltage Range
(P0621P)
0 to +10 V DC or peak AC
Input Voltage Range
(P0621N)
0 to -10 V DC or peak AC
Input Impedance
50
kW, nominal
DC Voltage Gain
3000 V/V
DC Voltage Gain
Accuracy
Better than 0.1% of full scale
Offset Voltage
Less than ±4 V
Output Noise
Less than 5 V rms*
Slew Rate
(10% to 90%, typical)
Greater than 350 V/µs
Large Signal
Bandwidth
(1% distortion)
DC to greater than 3.5 kHz
Small Signal
Bandwidth (-3dB)
DC to greater than 25 kHz
Automatic Power
Limit
Automatically limits the internal power
dissipation to protect the Models P0621P and
P0621N from overheating
Voltage Monitor
Ratio
1/3000th of the HV output signal
DC Accuracy
Better than 0.1% of full scale
Offset Voltage
Less than ±2 mV
Output Noise
Less than 20 mV*
Output Impedance
47
W
Current Monitor
Ratio
0.5 V/mA
DC Accuracy
Better than 1% if full scale
Offset Voltage
Less than ±10 mV
Output Noise
Less than 30 mV rms*
Bandwidth
DC to greater than 10 kHz
Output Impedance
47
W
Measurement and Power Solutions
TM
www.trekinc.com
*Measured using the true rms feature of the HP Model 34401A digital multimeter