Performance
Features (cont.)
General
Features
Measurement Range
Voltage Monitor Output
Dimensions
0 to ±10 kV DC or peak AC.
A BNC output provides a low-voltage
134 mm H x 432 mm W x 432 mm D
replica of the measured voltage.
(5.25² H x 17² W x 17²D).
Accuracy
Scale Factor
Better than ±0.1% of full scale
Weight
1/1000th of the measured
(as referred to the voltage monitor).
15 kg (33 lb).
voltage.
Output Noise
Speed of Response (10% to 90%)
Voltage Monitor Connector
Less than 20 mV rms (measured
Less than 200 ms for a 1 kV step
BNC connector.
using the true rms feature of the
change. Less than 5 ms for a
Ground Receptacle
Hewlett Packard Model 34401A
20 kV step change.
Green binding post.
digital multimeter).
Full Signal Bandwidth
Output Impedance
AC Line Cord Receptacle
DC to better than 50 Hz.
Less than 0.1 W.
Standard three-prong line cord
Stability
receptacle with an integral fuse
Voltage Display
Drift with Time
holder.
4½ digit LED display.
Less than 100 ppm/hour,
Range
Line Supply
noncumulative.
0 to ±10,000 V.
Factory set for one of three nominal
Drift with Temperature
Resolution
AC line voltages: 100 V, 115 V, or
Less than 200 ppm/°C.
1 V.
230 V AC [CE compliant], at
Zero Offset
48 to 63 Hz (specify when ordering).
Less than ±2 counts,
referred to the voltage monitor.
Operating Conditions
High-Voltage Ready LED
Sampling Rate
Temperature
A LED indicator illuminates when the
3 readings per second.
0 °C to 40 °C.
Model P0865 is ready to make high-
Relative Humidity
voltage measurements.
Power On/Off
To 90%, noncondensing.
A two-position rocker switch that turns
High Voltage On/Off
Probe-to-Surface Separation
on and off the main power.
A two-position toggle switch that turns
3 mm ±1 mm
(recommended).
on and off the high-voltage power
supply inside the instrument.
Zero Adjust
A 10-turn control to null offsets or
other zero errors which occur within
the system.
Model P0865
Probes
Optional Accessories
Certification
Item
Part No.
P0865 Electrostatic Voltmeter (100 V AC). . . . . . . . . . . . . . . . 13424F
Item
Part No.
P0865 Electrostatic Voltmeter (115 V AC) . . . . . . . . . . . . . . . 13424G
P0865 Electrostatic Voltmeter (230 V AC) . . . . . . . . . . . . . . . 13424K
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17157
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 17284
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 17285
Item
Part No.
Model 3
. . . 17218
Model 3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17181
Full Rack Mounting Kit
607RA Rack Mount Kit (19-inch) . . . . . . . . . . . . . . . . . . . C4062
Standard Resolution Probe
Model 3450 side-viewing
High Temperature Probes (up to 100 °C)
Model 3455ET end-viewing
Model 3453ST side-viewing
450EC Probe Extension Cable (from unit to probe)
460-1 Line Driver
TREK, INC. certifies that each Model P0865 is tested and calibrated
to specifications using measurement equipment traceable to the
National Institute of Standards and Technology or traceable to
consensus standards.
Model 10/40A Specifications
Model P0865 Specifications
Copyright ã 2011 TREK, INC.
All specifications are subject to change.
1147/DEC
Model P0865 Ordering Information
All specifications are with a Model 3450 probe at a probe-to-surface separation of 3 mm, ±1 mm.
TREK, INC. • 190 Walnut Street • Lockport, NY 14094 • USA • 800-FOR TREK
(fax) • www.trekinc.com • [email protected]
716-438-7555 • 716-201-1804