Performance
Features (cont.)
General
Features
Charged Plate Voltage Range
Dimensions
0 to ±55 V DC or peak AC,
122 mm W x 43 mm H x 153 mm D
nominal.
(4.8² W x1.7² H x 6² D).
Weight
Charged Plate Voltage
0.3 kg (0.6 lb).
Measurement Accuracy
0.2% of full scale.
Positive discharge tests,
The Model PD04002A operates
Large Signal Bandwidth (-3 dB)
Negative discharge tests, and
from a +24 V ±10% supply
DC to greater than 200 Hz.
Balance tests can be initialized
@ 0.21 amps.
from a remote location through
Small Signal Bandwidth (-3 dB)
these digital command signals.
DC to 2.5 kHz.
The voltage of the charged plate
is initialized to +55 V or -55 V for
Charged Plate Self-Discharge
discharge time testing or zero
Rate (no incident ion flow)
for balance testing.
Less than 2 V per minute at
55 V for relative humidity up to
85%.
Charged Plate Capacitance
20 pF ± 4 pF. (The capacitance
is independent of charged plate
connecting cable length.)
Unit Operating Conditions
Temperature
5 °C to 35 °C.
Relative Humidity
±
To 85%, noncondensing.
Relative Humidity
To 85%, noncondensing.
TREK, INC. certifies that each Model
Mounting
PD04002A Charged Plate Monitor is
Model PD04002A Monitor
tested and calibrated to specifications
Four (4) M4 mounting screws
using measurement equipment
traceable to the National Institute of
can be used to secure
Standards and Technology or traceable
the monitor assembly from the
to consensus standards. CE compliant.
bottom.
Caution: These instruments are
Charged Plate
designed to make electrostatic
One (1) M4 mounting screw
voltage measurements only. Do not
can be used to secure the
use this instrument to make
charged-plate monitor assembly
measurements of “hard” voltage
from the bottom.
sources or voltage sources which
can deliver currents high enough to
User Connections
cause harmful shocks or personal
Connection block and
injury.
M4 ground lug.
Power
Connection Block
Phoenix Contact Connector.
Connections include:
+24 V DC power input (+),
+24 V DC return (-),
Voltage Monitor (Hi),
Voltage Monitor (Lo),
Digital Reset (plate 0),
Digital Reset (plate +), and
Digital Reset (plate-)
Charged Plate Assembly
High-Temperature Charged
Input Connector
Plate (Trek Model PD04002AP)
BNC connector.
25 mm x 25 mm (1² x 1²).
Cable Length
3 meters.
Capacitance
Voltage Monitor Output V
20 pF 5 pF.
0 to +5 V DC represents a
Operating Range
charged plate voltage of
-60° to 160° C.
-55 V to +55 V (other scale
factors are available).
When the charged plate is at
-55 volts, the voltage at the
Voltage Monitor Output V HI is
0 V. When the charged plate is
at 0 volts, the voltage at the
voltage monitor is +2.5 volts.
When the charged plate is at
+55 volts, the voltage at the
voltage monitor is +5 volts.
Accuracy
0.2% of full scale.
Digital Command Terminals
Plate (O) discharges the floating
plate to zero volts.
Plate (+) precharges the floating
plate to +55 V.
Plate (-) precharges the floating
plate to -55V.
All the above digital command
inputs are at a 10 V DC High
Level. A minimum of 6 mA of
current sink capability is required
to pull the digital command inputs
to ground (0 V).
Certification