Measurement and power solutions, Model 30/20a specifications – Trek 30-20A High-Voltage Power Amplifier User Manual
Page 2

Features (cont.)
High-Voltage On/Off
Local
Individual push-button switches
Remote
TTL compatible input. TTL high (or open) turns
off high-voltage output. TTL low turns on high-
voltage output.
Current Limit/Trip
Switch selectable for limit or trip. Graduated 1-
turn panel potentiometer is used to adjust limit
or trip level from 0 to ±20 mA.
Out of Regulation
Status Indicator and
Connnector
Illuminates and TTL low is provided when unit
fails to produce required HV output such as
during current limit.
Limit/Trip Status
Indicator and
Connector
An indicator will illuminate and a BNC will
provide a TTL low when the high-voltage output
is disabled due to the output current exceeding
the current trip level, the detection of a high-
voltage supply fault, the removal of one of the
panels, or if the Model 30/20A is out of
regulation for greater than 500 ms.
Mechanical
Dimensions
103.9 cm H x 43 cm W x 87 cm D
(40.9” H x 17” W x 34” D) Depth dimension
includes wheels, handles and spacing for air
flow.
Weight
73 kg (160 lb) approximate
HV Connector
Caton high-voltage Connector
BNC Connectors
Amplifier Input, Voltage Monitor, Current Monitor,
Remote High Voltage ON/OFF, Out of Regulation
Status, Fault/Trip Status
Operating Conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative Humidity
To 75%, noncondensing
Altitude
To 1524 meters (5000 ft.)
Electrical
Line Voltage
Factory set for one of two ranges:
104 to 127 V AC or 180 to 250 V AC at 48 to
63 Hz (specify when ordering)
Power Consumption
1800 VA, maximum
AC Line Receptacle
Standard 3-prong with integral fuse holder
Supplied Accessories
Operators Manual
PN: 23343
Shorting BNC Cap
PN: B3060
HV Output Cable
PN: 43466
Line Cord, Fuses
Selected per geographic destination
Locking Wheel Kit
CN: 1K042
Model 30/20A Specifications
Performance
Output Voltage
Range
0 to ±30 kV DC or peak AC
Output Current
Range
0 to ±20 mA DC or peak AC
Input Voltage Range
0 to ±10 V DC or peak AC
Input Impedance
25 kW nominal (inverting/differential option
50 kW, nominal)
DC Voltage Gain
3000 V/V
DC Voltage Gain
Accuracy
Better than 0.1% of full scale
Offset Voltage
Less than ±4 V
Output Noise
Less than 1.5 V rms*
Slew Rate
(10% to 90%, typical)
Greater than 550 V/µs
Small Signal
Bandwidth (-3dB)
DC to greater than 30 kHz
Large Signal
Bandwidth
(2% distortion)
DC to greater than 2.5 kHz
Stability
Drift with Time
Less than 50 ppm/hr, noncumulative
Drift with
Temperature
-
Less than 100 ppm/°C
Voltage Monitor
Ratio
1 V / 3000 V
DC Accuracy
Better than 0.1% of full scale
DC Offset Voltage
Less than ±5 mV
Output Noise
Less than 20 mV rms*
Output Impedance
47 W
Current Monitor
Ratio
0.5 V/mA
DC Accuracy
Better than 2% of full scale
Offset Voltage
Less than ±10 mV
Output Noise
Less than 30 mV rms*
Bandwidth (-3dB)
DC to greater than 5 kHz
Output Impedance
47 W
Features
Settling Time (to 1%)
Less than 200 µs for a 0-30 kV step
Dynamic Adjustment
Graduated 1-turn panel potentiometer is used
to optimize the AC response for various load
parameters.