VEGA VEGAMIP R62 Receiving unit - Transistor User Manual
Page 41
l
1 x cable entry ½ NPT, 1 x blind stopper ½ NPT,
1 x cable entry ½ NPT
or:
l
1 x plug M12 x 1; 1 x blind stopper M20 x 1.5
Cable entry/plug - external housing
-
Housing socket
l
1 x cable entry M20 x 1.5 (cable: ø 6 … 12 mm)
-
Housing
l
1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5
or:
l
1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5
Spring-loaded terminals
for wire cross-section up to 1.5 mm² (AWG 16)
Connection cable to external housing
-
Structure
Three wires, double screening, weather-resistant
-
Material
PUR UL
94-V0
-
Wire cross-section
0.34 mm² (AWG 22)
-
Wire resistance
<
0.060 Ω/m (0.018 Ω/ft)
-
Capacitance - screen
≤
250 pF/m
-
Max. length
25 m (82 ft)
-
Min. bending radius at 25 °C/77 °F
25 mm (0.985 in)
-
Diameter approx.
7 mm (0.276 in)
-
Colour
Black
Voltage supply
Operating voltage
20 … 55 V DC
Power consumption
max. 0.8 W
Electrical protective measures
Protection rating
-
Instrument housing
IP
66/IP 67
-
External housing
IP
65
-
Housing socket - external housing
IP
68 (1 bar)
Overvoltage category
III
Protection class
II
Approvals
Instruments with approvals can have different technical data depending on the version.
VEGAMIP R
62 • - Transistor
41
9 Supplement
41911
-
EN
-120228