The quick start enables a quick setup with many applications. You

can find further information in the respective chapters of the operating

instructions manual.
1. Distance from the vessel wall > 200 mm

200 mm

(7.87")

Fig. 1: Distance of the antenna to the vessel wall/socket configuration

2. Keep the socket short, round off the socket end
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specifica-

tions on the type label.

2. Connect the instrument according to the following illustration

+

+

power supply

MODBUS

D0

D1

IS GND

USB

1

2

1

3 4 5

2

off on

( )

(

)

(-)

(-)

Fig. 2: Connection compartment
1 USB connection

2 Slide switch for integrated termination resistor (120 Ω)

For further information see chapter "Connecting to power supply".
1. Go via the indicating and adjustment module to the menu "Setup".

2. Select in the menu item "Medium" the medium of your applica-

tion, for example "Gravel/Pebble".

3. Select in the menu item "Application" the vessel, the application

and the vessel form, for example, silo.

Mounting

Connect electrically

Set parameters