Specifications, Ix2-rfaeva-2 – Olympus IX2 RFAEVA2 User Manual
Page 21
18
IX2-RFAEVA-2
SPECIFICATIONS
Item
Specification
Applicable microscope frames
· IX2 series (IX81/IX71/IX51)
· IX series (IX70/IX50)
Applicable reflected fluorescence
systems
· IX2 and IX series systems
(Reflected light illuminator unnecessary)
Total internal reflection illumination
system
IX2-RFAEVA
Straight (mercury burner)/Fiber light (laser) double illuminator
· Either illuminator can be selected using the illumination selector
knob.
· Interlock mechanism (Laser interlock/User interlock).
· Field iris diaphragm (Diameter adjustable between 1.7 and 13 mm)
· Filter slider
· Maximum FN 11.7
· Interlock key switch
· Laser emission indicators
Binocular tube shutter
· Laser interlock mechanism with the binocular tube shutter.
(Releasing the shutter engages the shutter in the laser light path.)
Specimen cover
· Laser interlock mechanism with the specimen replacement lid.
(Removing the specimen replacement lid engages the shutter in
the laser light path.)
Ar laser unit/power supply
FV5-LAAR
· 10 mW power, linear polarization
· Forced air-cooled Ar ion laser (488 nm)
· Connection with the fiber light illuminator: Polarization plane saving
fiber cable FV5-FUR (3 m)
· Power consumption: 115 V, 10 A (max.)
230 V, 10 A (max.)
Operating environment
· Indoor use
· Altitude: Max. 2000 m
· Temperature: 18 to 28°C (performance guaranteed), 10 to 35°C (laser
oscillation guaranteed)
· Relative humidity: 30% 80% (without condensation)
· Supply voltage fluctuations: ±10%
· Pollution degree: 2 (in accordance with IEC60664)
· Installation (overvoltage) category: II (in accordance with IEC60664)
Storage environment
· Temperature: –25 to 65°C
· Relative humidity: 10% to 90%
Safety
EN61010-1(1993 +A2 1995): Safety requirements for electrical equipment for measurement, control and laboratory
use.
EN60825-1(1994 +A2 2001): Safety of laser products.
Electromagnetic Compatibility
EN61326(1997 +A1 1998): Electrical equipment for measurement, control and laboratory use EMC requirements.
EN61326 defines two categories according to the location for use.
Class A: Equipment suitable for use in establishments other than domestic, and those directly connected
to a low voltage power supply network which supplies buildings used for domestic purposes.
Class B: Equipment for use in domestic establishments, and in establishments directly connected to a low
voltage power supply network which supplies buildings used for domestic purposes.
This system is applied Class A. Some interference may occur if this system is used in domestic location.