VEGA Capacitive electrodes EK Profibus PA User Manual
Page 37
Capacitive electrodes EK - Profibus PA
37
First of all, click to "Profibus DP for PA
sensors". Then choose the Profibus card
which is installed in your PC and click to
OK
.
Then restart VVO. If you are connected via
VEGACONNECT 3, choose the option
Stan-
dard
(for connection via VEGACONNECT 3
you require VVO version 2.80 or higher).
Sensor configuration
If you start VVO as described above, i.e. if
you have connection to the Profibus, you will
see the following presentation on your
screen. Click to
Configuration
, then to
Measuring system
.
Display Diagnostics Instrument data
Configuration Services Quit Help
Measuring system
Measurement loop >
Program
>
Language
User access
Communication
If you have already connected several VEGA
sensors to the bus, you can choose the
sensor you want to configure - in this case,
the capacitive electrode. The serial number
and the bus address of the sensor will be
displayed to ensure a clear identification.
Through VEGACONNECT 3 you can access
only one sensor.
Configuration
Setup
• Switch on the power supply of the
connected sensor.
• Start the adjustment software VVO (VEGA
visual Operating) on your PC.
• In the entrance screen, you choose with
the arrow keys or the mouse the item
Planning and click to OK. You should only
choose Planning if you are authorised to
modify the instrument parameters.
Otherwise choose Operator or
Maintenance. In the window "User
identification", you are asked for the name
and the password.
• For setup (planning) you enter under
name: VEGA and under password also:
VEGA. It does not matter if you use capital
or small letters.
VVO recognises automatically the type of the
connected sensor and immediately
afterwards indicates with which sensor a
connection exists.
The following adjustment steps are
described in their sequence and should be
carried out in the same order during the initial
setup.
For further information, see the operating
instructions of the adjustment program VEGA
Visual Operating (VVO).
If you cannot get a connection to the
instrument, click to
Configuration
, choose
Program
and click to Communication.