VEGA VEGASON 71 - 2 … 75 - 2 User Manual
Page 54
Supplement
54
71 - 2 … 75 - 2
Mode-no.
Mode description
Factory preset
Sensor optimization
1 - or 2 -
1
2
3
Working range
07
-
start .......................................................
0.00 m
08
-
end
71 - 2 ...................................
6.00 m
72 - 2 ...................................
12.00 m
73 - 2 ...................................
12.00 m
74 - 2 ...................................
24.00 m
75 - 2 ...................................
36.00 m
Product type
09
-
liquid / solid
for VEGASON 72 ..................................
1 (liquid)
for VEGASON 73 … 75 ........................
0 (solid)
Solid
10
-
angle of repose in degrees ...................
0 degrees
11
-
particle size ...........................................
11 mm
Liquid
10
-
vessel top flat / dished ..........................
1 (flat)
Multiple false echo
12
-
reduction in Volt ....................................
0.00
13
-
optimization ...........................................
0.00
False echo learn
15
-
learn distance in m ................................
2.00
16
-
delete learn cycle ..................................
0 (off)
17
-
activate learn cycle ...............................
0 (off)
18
Profile interpolation ....................................
0 (off)
19
Rate of product level change in cm / sec. ..
4.00
Max. gain
21
-
limitation in Volt .....................................
0.03
22
-
optimization in Volt ................................
0.03
23
Measuring window .....................................
1 (on)
24
Fault signal ................................................
0 (off)
25
Real distance in m .....................................
2.00
27
Basic adjustment .......................................
0 (off)
Date
Name
Protocol of the sensor optimization