VEGA VEGASON 71 - 2 … 75 - 2 User Manual

Page 22

background image

Adjustment at differential measurement

22

71 - 2 … 75 - 2

Differential measurement

A differential measurement can be obtained with
the measuring system VEGASON 71 - 2 … 75 - 2.

The difference is formed out of the values of chan-
nel 2 minus channel 1. If possible, mount the
sensors at the same height.

The adjustment is made for both measuring chan-
nels together in mode range no. 30 … 32.

It is possible to adjust channels 1 and 2 separately
as described under level measurement.

The display coordination is defined as follows in
mode 0 - 40:

0 - 4 0

1 1

0 = display 2 switched off
1 = display 2 indicates level

of channel 2

2 = display 2 indicates difference

0 = display 1 switched off
1 = display 1 indicates level

of channel 1

2 = display 1 indicates difference

The pre-adjustment 1 for mode 0 - 30 should not be
changed.

Mode description

Mode 0 - 30 = 1

mean adjustment in m without product level change, i.e.
- with this adjustment procedure various differences (levels, gauges etc.) are

programmed relating to the values 0 % and 100 %

- already available levels or water gauges are unimportant

Selection of adjustment procedure

MODE 0 - 30

MODE 0 - 40

Programming example
for

- differential level on display 1
- level 2 (downstream water) on display 2 .......

• select mode 0 - 40 in MODEFIELD
• program figure 21 in PARAMETER FIELD
• then store with STO-button