beautypg.com

VEGA VEGASON 71 - 2 … 75 - 2 User Manual

Page 17

background image

Start-up

71 - 2 … 75 - 2

17

Mode listing 0

general parameters

Display

Multiplier

function

0 - 23

- related for 0 % ........................................

0000

9999

0 - 24

- related for 100 % ....................................

0000

1000

9999

0 - 25

Measuring unit
- level percentage in m .............................

1

p. 29

- level percentage in % .............................

2

- volume percentage (scaled) ...................

3

0 - 26

Decimal point ...........................................

0.000

000.0

0000

0 - 27

Integration time ........................................

0

900

Lineari-

0 - 28

Linearization curves

zation

- linear ......................................................

1

- cylindrical tank ........................................

2

- not used .................................................

3

p. 24

- individually programmable curve 4 .........

4

- individually programmable curve 5 .........

5

- individually programmable curve 6 .........

6

0 - 29

not used .....................................................

– – – –

Adjustment

0 - 30

Selection of adjustment procedure

differential

- in m without level change .......................

1

measure-

- percent correction ..................................

2

ment

0 - 31

Adjustment min. difference
- in m .........................................................

sensor specific

p. 22

- percent correction ..................................

0.0

80.0

0 - 32

Adjustment max. difference
- in m .........................................................

sensor specific

- percent correction ..................................

100.0

20.0

Display

Multiplier

function

0 - 33

- related to 0 % .........................................

0000

9999

0 - 34

- related to 100 % .....................................

0000

1000

9999

0 - 35

Measuring unit
- level percentage in m .............................

1

p. 29

- level percentage in % .............................

2

- volume percentage (scaled) ...................

3

0 - 36

Decimal point ...........................................

0.000

000.0

0000

0 - 37

39

not used .....................................................

– – – –

Display

0 - 40

Display coordination

1 1

0

= inactive

p. 22

Display 2

1

= level indication

(channel specific)

Display 1

2

= differential indication

0 - 41

u.

42

not used .....................................................

– – – –

Function

Mode-no.

Mode description

Parameter

Page

(

= factory preset)

(continuation)