VEGA VEGASON 71 - 2 … 75 - 2 User Manual
Page 17
Start-up
71 - 2 … 75 - 2
17
Mode listing 0
general parameters
Display
Multiplier
function
0 - 23
- related for 0 % ........................................
0000
…
9999
0 - 24
- related for 100 % ....................................
0000
…
1000
…
9999
0 - 25
Measuring unit
- level percentage in m .............................
1
p. 29
- level percentage in % .............................
2
- volume percentage (scaled) ...................
3
0 - 26
Decimal point ...........................................
0.000
…
000.0
…
0000
0 - 27
Integration time ........................................
0
…
900
Lineari-
0 - 28
Linearization curves
zation
- linear ......................................................
1
- cylindrical tank ........................................
2
- not used .................................................
3
p. 24
- individually programmable curve 4 .........
4
- individually programmable curve 5 .........
5
- individually programmable curve 6 .........
6
0 - 29
not used .....................................................
– – – –
Adjustment
0 - 30
Selection of adjustment procedure
differential
- in m without level change .......................
1
measure-
- percent correction ..................................
2
ment
0 - 31
Adjustment min. difference
- in m .........................................................
sensor specific
p. 22
- percent correction ..................................
0.0
…
80.0
0 - 32
Adjustment max. difference
- in m .........................................................
sensor specific
- percent correction ..................................
100.0
…
20.0
Display
Multiplier
function
0 - 33
- related to 0 % .........................................
0000
…
9999
0 - 34
- related to 100 % .....................................
0000
…
1000
…
9999
0 - 35
Measuring unit
- level percentage in m .............................
1
p. 29
- level percentage in % .............................
2
- volume percentage (scaled) ...................
3
0 - 36
Decimal point ...........................................
0.000
…
000.0
…
0000
0 - 37
…
39
not used .....................................................
– – – –
Display
0 - 40
Display coordination
1 1
0
= inactive
p. 22
Display 2
1
= level indication
(channel specific)
Display 1
2
= differential indication
0 - 41
u.
42
not used .....................................................
– – – –
Function
Mode-no.
Mode description
Parameter
Page
(
= factory preset)
(continuation)