The most important adjustment steps – VEGA VEGABAR 40 HART User Manual
Page 31
VEGABAR 40 (HART
®
)
31
Setup
The most important adjustment steps
On the following four pages you see a menu
schematic of the HART
®
handheld in conjunc-
tion with VEGABAR process pressure trans-
mitters. The most important adjustment steps
are marked in the menu schematic with the
letters A … E. If you are not familiar with
HART
®
handheld, please note: For the pa-
rameter adjustment, first push the "
ENTER“
key. The adjustment is saved in the hand-
held, but not in the sensor itself.
After you have pushed "
ENTER“, you have to
push "
SEND“ (here in the example for the
min. adjustment).
After pushing "
SEND“, a warning is displayed
that you are about to change device output
and for safety reasons, you should switch
your system to manual operation.
Push "
OK“ and the adjustment is now trans-
ferred to the sensor. After a short moment,
you are asked to switch the system from
manual to automatic operation. Confirm with
"
OK“.
You now see the adjustment that was made.
4.2
(5.2)
Generic: SENSOR
PV URV
100.0 mbar
90.3
mbar
HELP DEL
ESC ENTER
Generic: SENSOR
1 PV
LRV bar
2 PV
URV bar
HELP SEND
HOME
Generic: SENSOR
– WARNING –
Pressing "OK“ will
change device output.
Put loop in manual.
ABORT OK
Generic: SENSOR
– WARNING –
Return control loop
to automatic control
OK
Generic: SENSOR
1 PV
LRV bar
2 PV
URV bar
HELP