beautypg.com

Ohaus DEFENDER 5000 BENCH SCALES Manual en User Manual

Page 17

background image

EN-15

5000 Series Indicators

3.2 Menu Navigation

3.3 Calibration Menu

When CAL is displayed, press the

Yes button to accept the Calibration menu

selection. Press the

No button to advance to the desired calibration menu item.

Three calibration processes are available: Zero Calibration, Span Calibration and
Linearity Calibration. Default settings are

bold.

Zero Perform
Span Perform
Linearity Perform
Cal Test Perform
Geographic
Adjustment Set 00…

Set 12…Set 31

End Calibration Exit CALIBRATE menu

NOTES:
1. Make sure that appropriate calibration masses are available before
beginning calibration.
2. Make sure that the scale base is level and stable during the entire
calibration process.
3. Calibration is unavailable with LFT set to ON.
4. Allow the Indicator to warm up for approximately 5 minutes after
stabilizing to room temperature.
5. To abort calibration, press the

Exit button anytime during the calibration

process.
6. When any selection within the GMP menu is enabled, calibration results
are automatically printed.

Enter the menu by pressing the

TARE

Menu button until MENU is displayed. When the button is released,

the Legal for Trade status is displayed, followed by the first menu. Press the

No or Back button to move to a

different menu. Press the

Yes button to enter the menu. Once in the menu, press the Yes button to view the

menu item setting or press the

No or Back button to move to the next menu item. When viewing the setting,

press the

Yes button to accept the setting, or press the No or Back button to change the setting. Once all

settings have been made, press the

Exit button to return to the current application mode.

For menu items with numeric settings such as Capacity, the current setting is displayed with all digits flashing.
Press the

No button to begin editing.

The first digit is displayed flashing.

Press the

No button to increment the digit or press the Yes button to accept the digit and move to the next digit.

Repeat this process for all digits.

Press the

Yes button when the last digit has been set.

The new setting is displayed with all digits flashing. Press the

Yes button to accept the setting or press the No

button to resume editing.

This method also applies to setting Checkweigh under and over targets.

For End menu items, pressing the

Yes button advances to the next menu, while pressing the No button returns

to the top of the current menu.