Ohaus DEFENDER 3000 METAL WEIGHING INDICATOR Manual en User Manual
Page 33

EN-31
3000 Series Indicators
4.6.3 Display-Hold
Use this mode to capture and store the first stable display value.
To enter the Display hold mode, press and hold the
Mode button until HOLd is displayed.
When the
Mode button is released, the Hold indicator will be turned on.
Place the item on the scale, The first stable value will be held on the display and the Hold icon will
blink. When a successive weight is added to the scale the display will show the new value.
Even after removing the items,the weight will be held on the display.
To clear the held value, press
Mode button.
NOTE: Press and hold the
Mode button to enter other modes.
4.6.4 Check Weighing
Use this mode to determine if the weight of a sample is within Pre-set limits.
Checkweighing Limits
When the FUNCTION
Mode button is released, Clr.rEF is displayed.
Using or Recalling Stored Check Weighing Limits
Press the No button to recall the stored limits and proceed to checkweigh operation
Clearing Stored Check Weighing Limits
Press the Yes button to clear the stored limits.
NOTE: Press the FUNCTION
Mode button successively to temporarily display the Under and Over
Limit values.
Editing the Under Setting
The display shows SEt.LO. Press the Yes button to edit setting..
Settings:
-99999 to 999999
Refer to Menu Navigation Section 3.2 to enter settings.
NOTE: The first digit will be used to show a negative value. Adjust the readability setting as needed
to allow an extra digit taken up by the negative sign
to