3000 Series Indicators

Reset No, Yes
Legal for Trade

Off, On

Calibration Unit kg, lb
Capacity

5…20000

Graduation

0.0005…20

Power On Unit Auto, kg, lb, g, oz
Zero Range 2%, 100%
Retain Zero Data Off, On
End Setup Exit SETUP menu

3.4 Setup Menu

When the Indicator is used for the first time, enter this menu to set the
Capacity and Graduation.

3.4.1 Reset

Reset the Setup menu to the factory defaults.
No

= not reset.

Yes = reset.

3.4.4 Capacity

Set the scale capacity from 5 to 20000. Refer to the Setup Table 3.3 for available settings.

3.4.2 Legal for Trade

Set the legal for trade status.
OFF = off
ON = on

Turning on the “LFT” menu setting has the following effects:
• Zero-range is set and locked on “2”.
• Auto Zero Tracking is set and locked on 0.5d
• The lb:oz unit is not available as a power-on setting.

3.4.3 Calibration Unit

Set the unit during calibration.
CAL UN kg = Calibrate using kg weights
CAL UN lb = Calibrate using pound weights