Ohaus DEFENDER 3000 METAL WEIGHING INDICATOR Manual en User Manual
Page 23
EN-21
3000 Series Indicators
3.5 Readout Menu
Enter this menu to customize display functionality.
Reset No, Yes
Stable 0.5d, 1d, 2d, 3d
Filter Level Lo, Med, Hi
Auto-Zero Tracking Off, 0.5d, 1d, 3d
Sleep (T32ME)
Off, On
Light (T32ME) Lo, Med, Hi
Backlight (T32MC) Off, On, Auto
Auto Off Timer Off, Set 1, Set 2, Set 5
Expand Off
Off, On
End Readout
Exit READOUT menu
3.5.1 Reset
Set the Readout menu to factory default settings.
No
= not reset
Yes = reset
3.5.4 Auto-Zero Tracking
Set the automatic zero tracking functionality.
OFF
= disabled.
0.5 d = the display will maintain zero until a drift of 0.5 divisions per second has been
exceeded.
1 d = the display will maintain zero until a drift of 1 division per second has been
exceeded.
3 d = the display will maintain zero until a drift of 3 divisions per second has been
exceeded.
NOTE: When the LFT menu item is set to ON, the selections are limited to 0.5d. The setting is locked when
the hardware lock switch is set to the ON position.
3.5.3 Filter
Set the amount of signal filtering.
LO = less stability, faster stabilization time
MEd = normal stability, stabilization time
HI = greater stability, slower stabilization time
3.5.2 Stable
Set the amount the reading can vary while the stability symbol remains on.
0.5 d
= 0.5 divisions
1 d = 1 division
2 d = 2 divisions
3 d = 3 divisions