Operation – American Magnetics 4Q06250PS-430 Integrated Power Supply System User Manual
Page 52

34
Rev. 5
Operation
Shift Key Commands : Ramp Rate
is installed), attempting to use the fine adjust knob for PSw P/S
ramp rate will produce one beep
1
.
Pressing
SHIFT
and then
RAMP RATE
will access the ramp rate
menu. The numeric and
ENTER
keys (or the fine adjust knob) are
used to set the segment 1 ramp rate to a value of 0.2.
The right arrow key is pressed once to access the segment 1 range
display. The numeric and
ENTER
keys (or fine adjust knob) are
used to set the segment 1 current range upper bound to a value of
55.
Pressing the right arrow key accesses the next (second) segment
ramp-rate display. The segment 2 ramp rate is set to a value of 0.1.
The right arrow key is pressed once to access the segment 2 range
display. The segment 2 current range upper bound is set to a value
of 58.
5. Since the PSw P/S ramp rate is active in that scenario, and not the segmented
ramp rate.
1. Since the standard segmented ramp rate is active in that scenario, and not the PSw
P/S ramp rate.
+50.00 A - Seg.1 Ramp Rate (A/sec)
+0.50 Vs
±0.2000
+50.00 A - Seg.1 Range (A)
+0.50 Vs
0.0 to ±55.0
+50.00 A - Seg.2 Ramp Rate (A/sec)
+0.50 Vs
±0.1000
+50.00 A - Seg.2 Range (A)
+0.50 Vs
±55.0 to ±58.0