Setup testing the scales – ACU-RITE Micro-Line M User Manual
Page 14

MICRO-LINE
Reference Manual
11
SETUP
Testing the Scales
Follow these steps to confirm that the scale has been installed properly.
The Counting Test will confirm the scale’s electrical operation, and the
Repeatability Test will check the installation integrity.
Counting Test
•
Move the table along each axis, one at a time. Check that the 
readout correctly displays the table’s movement. If it doesn’t, then 
the scale resolution you selected may not be accurate. Refer to 
Scale Resolution on pg. 9 for instructions.
Repeatability Test
•
Locate a magnetic base and dial indicator on the end of the scale. 
Zero the readout and the indicator.
•
Move the axis through the full travel and return the dial to “0.” 
The readout should also read 0 (±.0005” for 10µm scales; 
±0.0002” for 5µm scales). If it doesn’t, then the scale cable may 
be loose, or the scale, mounting bracket, or reading head may 
need to be tightened or realigned.
•
Repeat these steps for each scale.