Adjusting the Diopter (Zoom Confocality Adjustment)

(Fig. 3)

# Confirm that the eyepiece clamping knobs @ are tightened firmly

before proceeding to the adjustment.

Eyepieces without eyepiece micrometer disk

1. Turn the diopter adjustment rings ² of the left and right eyepieces to

positions “0”. (This adjustment is not possible with eyepieces without
helicoids.)

2. Place an easy-to-observe specimen on the stage plate.
3. Set the zoom adjustment knob ³ for the lowest magnification and rotate

the focus adjustment knob | to bring the specimen into focus.

4. Set the zoom adjustment knob ³ for the highest magnification and

rotate the focus adjustment knob | to bring the specimen into focus.

5. Set the zoom adjustment knob ³ for the lowest magnification and rotate

the diopter adjustment rings ² of the left and right eyepieces, instead of
the focus adjustment knob, to bring the specimen into focus.

}Set the zoom adjustment knob ³ for the highest magnification again

and check the image focusing. The diopter adjustment is complete if the
image in accurate focus. If not, repeat steps 3 to 5 above.

Eyepieces with eyepiece micrometer disk

1. Look through the eyepiece with eyepiece micrometer disk and rotate its

diopter adjustment ring ² to bring the micrometer scale into focus.

2. Place an easy-to-observe specimen on the stage plate.
3. Set the zoom adjustment knob ³ for low magnification, look through the

eyepiece with eyepiece micrometer disk and rotate the focus adjustment
knob | to bring the specimen into focus.

4. Set the zoom adjustment knob ³ for the highest magnification, look

through the eyepiece with micrometer disk and rotate the focus adjust-
ment knob | to bring the specimen into focus.

5. Set the zoom adjustment knob ³ for the lowest magnification and rotate

the diopter adjustment ring of the eyepiece without eyepiece micrometer
disk, instead of the focus adjustment knob, to bring the specimen into
focus.

}Set the zoom adjustment knob ³ for the highest magnification again

and check the image focusing. The diopter adjustment is complete if the
image in accurate focus. If not, repeat steps 3 to 5 above.

}Note the left and right diopter scale values so that it can be quickly

duplicated in future observations.

Diopter scale of the 10X eyepieces

}The valid range of the diopter scale is between -8 and +5 but a slight

margin is added to it. Therefore, the diopter value may exceed +5 or -8
when it is adjusted to the maximum. In this case, whether the diopter
value is over +8 or under -5 can be identified by the length of the
eyepiece.

Fig. 3

³

|

²

1

- 8

+ 5