4.2.3 Measuring Ranges
Values of measured quantities for which errors are defined.
Voltage
70 .. 120% of nominal
Current
5 .. 120% of nominal
Frequency
45 .. 66 Hz, 360 .. 440 Hz (Model 0440)
Crest values of voltage and current must remain within
168% of nominal maximum rms values
4.2.4 Accuracy
Voltage
0.4% of reading ±0.1% of range
1% of range maximum for Model 0440
Current
0.4% of reading ±0.1% of range
1% of range maximum for Model 0440
Frequency (not 0340)
0.15% of mid frequency
1% of mid frequency for Model 0440
Temperature coefficient
0.013%/°C typical
Response time to step input
1.5 seconds approx.
Screen update time
0.5 second approx.
4.2.5 Reference conditions of influence quantities
Values that quantities which affect measurement errors to a minor degree
have to be for the intrinsic (headline) errors for measured quantities to apply.
Ambient temperature
23°C
Input frequency
50 or 60 Hz 2%
Input waveform Sinusoidal
(distortion factor 0.005)
Auxiliary supply voltage Nominal
1%
Auxiliary supply frequency Nominal
1%
Auxiliary supply distortion factor
0.05
Magnetic field of external origin
Terrestrial flux
4.2.6 EMC Standards
EMC Immunity EN61326 for Industrial Locations to performance criterion A
EMC Emissions EN61326 to Class B - Domestic