Electrical specification – Tektronix 1705A User Manual
Page 23

Introduction
1- 4
1705A Spectrum Monitor
ELECTRICAL SPECIFICATION
Table 1- 1: Spectrum Display
Characteristic
Performance
Requirements
Supplemental Information
Check
Step
Frequency Range
L--- Band
70 MHz
950 to 1800 MHz
45 to 100 MHz
F--- type connector
Bnc connector
Frequency Span
L--- Band
Full
10 MHz/Division
1 MHz/Division
100 kHz/Division
70 MHz
Full
1 MHz/Division
100 kHz/Division
10 Horiz. Div. Equals:
1000 MHz
100 MHz
10 MHz
1 MHz
10 Horiz. Div. Equals:
50 MHz
10 MHz
1 MHz
11
5
Span/Div Accuracy
Typically 0.5 minor Division.
Brightup offset by 1 Division in FULL
SPAN/DIV should still come on screen
in the next magnified position.
4, 5, 10, 11
11
Flatness
L--- Band ±5 dB.
70 MHz ±2 dB.
± from center (1400 MHz).
± from center (70 MHz).
12,
7
Maximum Signal Input
L--- Band: --- 30 dBm,
75
Ω. 70 MHz:
--- 20 dBm, 75
Ω..
Minimum Signal Input
--- 80 dBm.
Relative Amplitude Accuracy
L--- Band (only)
±3 dB/100 MHz.
Typically ±1 dB /100 MHz.
12
Sweep Length
¶12 Divisions all SPANS/DIV settings. 4
Sweep Speed
Typically 20 --- 200 ms.
9
Positioning Range
Vertical
Horizontal
+ and --- 3 Divisions.
+ and --- 2 Divisions.
13
Displayed Frequencies in FULL
SPAN/DIV
L--- BAND (900 --- 1900 MHz)
70 MHz (45 --- 100 MHz)
Left Grat.
Mid Grat.
Right Grat.
Edge
Line
Edge
900
1400
1900
45
70
95
11
5
Frequency Readout
Center Frequency shown by time
sharing graphic readout. Exact posi-
tion on the trace of the center frequen-
cy shown by a caret in all spans except
FULL.
11