Checks and Adjustments

5- 22

1705A Spectrum Monitor

7.

Adjust Readout Position

Display the DAC test pattern by completing parts a. through c.

a.

Press both the 1705A SPAN/DIV buttons simultaneously to display the
Main Menu.

b.

Press either SPAN/DIV button to move the cursor to TEST, then press
the INPUT button to display the TEST menu.

c.

Press either SPAN/DIV button to move the cursor to DAC, then press
the INPUT button to display the DAC test pattern.

d. ADJUST

-- R27 and R29 (HORIZ and VERT POSN READOUT) for a

display crossing at the crt center.

e.

Exit the DAC test pattern by first pressing the INPUT button to display
the TEST menu.

f.

Press either SPAN/DIV button to move the cursor to EXIT, then press
the INPUT button twice to return to a normal display.

8.

Adjust 360/119.8 MHz L.O.

NOTE. Use an insulated tool for the following adjustments.

a.

Adjust the VERTICAL POSITION control so that the crt trace is on the
--70 graticule reference line.

b.

Connect the signal generator to J11 (480 MHz IF Input). Set the output
Frequency for 489.7 MHz and the Amplitude to --40 dBm.

c.

Select 300 kHz RESOLUTION.

d.

Set C55 to full mesh, C69 to 1/8 mesh, and L4 to the top of the coil
(three slug threads visible).

e.

ADJUST

-- FL1 for maximum signal (three screws beginning at the

outside adjustment screw). If no signal is present, adjust L4 down a few
turns and then adjust FL1.

f.

ADJUST

-- C55 for maximum signal. (C55 should peak at about 1/8 to

1/2 mesh.)

g.

ADJUST

-- FL1, center and inside screws (do not readjust the outside

screw), for maximum signal. (Less than one turn should be required.)

h. ADJUST

-- FL2 for maximum signal (three screws beginning at the back

screw).