Checks and Adjustments
5- 18
1705A Spectrum Monitor
15.
Adjust Span/Div Magnifier Range (R38 and R37)
16.
Adjust Intensified Zone Range (R61 and R36)
17.
Adjust Vertical Gain (R176, R239, R240, and R232)
18.
Adjust 2 dB/Div Position (R140)
19.
Adjust 590 MHz Notch (C69)
Figure 5-7 shows the 1705A Spectrum Monitor internal adjustment locations.
PRELIMINARY SETUP
Connect the 1705A ac power cord to the variable autotransformer. Turn the
1705A power on. Set the front-panel controls to start this procedure as
shown in Table 5--2.
Table 5- 2: Preliminary Control Settings
POWER
ON
INTENSITY
Set to Preference
FOCUS
SCALE
VERTICAL POSITION
Set later
HORIZONTAL POSITION
INPUT
70 MHz
RESOLUTION
300 kHz
VIDEO
ON
SPAN/DIV
FULL
CENTER FREQUENCY
anywhere
SWEEP SPEED
y
A1 POWER SUPPLY BOARD
(Refer to Figure 5-7 for adjustment locations.)
1.
Adjust +5 V
a.
Connect the DMM negative lead to TP1 (GND) and the positive lead to
W1 (+5V).
b.
ADJUST -- R99 (+5V ADJ) for +5.0 V
ᐔ0.5V.
Long Form Procedure