Checks and Adjustments
5- 12
1705A Spectrum Monitor
10. Check L--Band Linearity
REQUIREMENT
-- One marker per Division
±1 minor Division.
a.
Connect the output of the leveled sine wave generator, through the comb
generator, to the 1705A L--BAND INPUT. See Figure 5-4.
SG503
Comb
Generator
VARIABLE
AUTOTRANSFORMER
L--BAND
INPUT
1705A
Figure 5-4: Equipment connections for L- Band checkout
b.
Set the leveled sine wave generator Frequency to 100 MHz, Amplitude
Multiplier to X1, and the Amplitude to 1.5 (8 dBm).
c.
Set the 1705A INPUT to L--BAND.
d.
Set the SPAN/DIV to FULL.
e.
CHECK
-- for 10 frequency markers. (Ignore any sweep start marker.)
f.
CHECK
– that each marker is within one minor Division of a major
graticule line when the 2
nd
marker is on the first graticule line in from
the left. See Figure 5-5.