Oregon Scientific BPW810 User Manual

Page 15

background image

EN

14

Problem

Symptom

Check This

Remedy

Error
message

Err
shows
on the
display.

A measure-

ment error
occurred

Relax for a
moment and
then measure
again

Err 1, 2,
or 3
shows
on the
display

The wrist
cuff is not
secure

Refasten the
cuff and then
measure again

Err 4
shows
on the
display

The monitor
detected
motion while
measuring

Movement can
affect the
measurement.

Relax for a
moment and
then measure
again

Err 5
shows
on the
display

Pressure is
over 37.3
kPa (280
mmHg)

Relax for a
moment and
then measure
again

Err 6
shows
on the
display

Deflation
period was
too long

Movement can
affect the
measurement.

Relax for a
moment and
then measure
again

Problem

Symptom

Check This

Remedy

Error
message

Err 7 or 8
shows
on the
display

A calibration
error
occurred

Retake the
measurement. If
the problem
persists, contact
the retailer or our
customer service
department for
further assistance.
Refer to the
warranty for
contact
information and
return instructions

Settings
are
wrong

Date and
time are
incorrect

The clock
was not set or
reset after
installing new
batteries

Reset the clock

Measure-
ment unit
(kPa or
mmHg)
are
incorrect

The
measurement
unit was not
set or reset
after installing
new batteries

Reset the
measurement
unit