ENTER READING CONFIGURATION MENU:
Press
a
1) Press
a, if necessary, until CNFG prompt appears.
Press
d
2) Display advances to
INPT Input Menu.
Press
a
3) Display advances to
RDG Reading Configuration Menu.
Press
d
4) Display advances to
DEC Decimal Point.
DECIMAL POINT SUBMENU:
Press
d
5) Display flashes previous selection for Decimal location.
Press
b
6) Scroll though the available selections and choose Decimal
location:
FFFF, FFF.F, FF.FF or F.FFF
Press
d
7) Display shows
STRD stored message momentarily only, if
changes were made, otherwise press
a to advance to LOAD
Known/Unknown Loads Submenu.
Decimal Point is passive.
KNOWN/UNKNOWN LOADS SUBMENU:
Press
d
8) Display flashes previous selection of
ENBL Enable or DSBL
Disable.
Press
b
9) Scroll though the available selection of
ENBL or DSBL
(flashing).
Press
d
10) Display shows
STRD stored message momentarily and then
advances to
L.PNT Linearization Points Submenu.
If
ENBL Known Loads scaling method was selected, calculate
the input values to the instrument based on the actual signal
being received. If
DSBL Without Known Loads scaling method
was selected, calculate input values to the instrument based on
the transducer specification.
LINEARIZATION POINTS SUBMENU:
Press
d
11) Display flashes previous selection of Linearization Points
Submenu.
Press
b
12) Scroll though the available selections:
0002, 0003, 0004,
0006, 0006, 0007, 0008, 0009, 0010 - up to 10
Linearization Points can be selected. Default is
0002.
If display flashes
NONE, your instrument has only 2 linearization
points.
Press
d
13) Display shows
STRD stored message momentarily only, if a
change was made, otherwise press
a to advance to the FLTR
Filter Constant Submenu.
Linearization Points allow users to customize the Transducer curve.
20